	C, H, AND N ANALYZER
	Analysis of a Compound for Carbon, Hydrogen and Nitrogen

Principle
	The Model 240B Elemental Analyzer accurately determines the carbon, hydrogen and nitrogen content of organic compounds (CO2, H2O and nitrogen). Combustion occurs with pure oxygen under static conditions.  The combustion products are then analyzed automatically in a self-integrating, steady state, thermal conductivity analyzer which eliminates the tedious weighing of absorption traps associated with classical gravimetric analysis.

Procedure

	A. Run a blank and conditioning sample (acetanilide) following        the attached directions.

	B. Run a blank again then go to C.

	C. Calibrate the Perkin Elmer Elemental Analyzer with acetanilide     following the attached directions (sample size 1-3 mg).  This    must be done at the beginning of each laboratory period.           Determine K values.

	D. Obtain an unknown and determine the elemental composition of       this compoundx (sample size 1-3 mg).  Run 2-3 trials of the        unknown if time permits.






HHT/dat